TEMPERATURE
Temperature Sensor: Resistance Temperature Detector (RTD)
Temperature Range: -40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution: 0.01 °C
Temperature Response
Time in Air:
t63 - 8 minutes to 63% of change
t90 - 68 minutes to 90% of change
Calibrated Accuracy: ±0.5 °C (0 °C to 55 °C)
HUMIDITY
Humidity Sensor: Capacitive Polymer
Humidity Range: 0 %RH to 100 %RH (non-condensing)
Humidity Resolution: 0.1 %RH
Humidity Response
Time in Air:
t63 - 30 seconds to 63% of change
t90 - 2 minutes to 90% of change
Calibrated Accuracy: ±3.0 %RH maximum
Start Modes:
Software programmable immediate start or delay start up to 24 months in advance
Real Time Recording: May be used with PC to monitor and record data in real time
Memory: 16,350 readings per channel
Password Protection: Yes
Reading Rate: 1 reading every second up to 1 reading every 24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Memory Wrap: Yes
Data Format: Date and time stamped °C, °F, K, °R ; %RH, mg/mL, Dew Point
Battery Life: 2 years typical at 25 °C (15 minute reading rate)
Battery Type: 3.6V lithium battery included, user replaceable
Time Accuracy: ±1 minute/month at 25 °C
Required Interface Package: IFC400 or IFC406
Computer Interface: USB (interface cable required); 125,000 baud
Operating System
Compatibility: Windows XP SP3/Vista/Windows 7/Windows 8
MadgeTech Software
Compatibility: Standard Software version 4.1.0.2 or later
Operating Environment: -40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH (non-condensing)
Dimensions: 1.7 in x 0.97 in x 0.97 in (42 mm x 24.6 mm x 24.6 mm)
Weight: 2.3 oz (65 g)
Material: 316 Stainless Steel/PEEK
Approvals: CE; Intrinsically Safe for Class I, Div 1, groups ABCD; Non-incendive for Class I, Division 2, groups ABCD Hazardous (Classified) Locations
*BATTERY WARNING: FIRE, EXPLOSION AND SEVERE BURN HAZARD. DO NOT RECHARGE,
DISASSEMBLE, HEAT ABOVE 100 °C (212 °F), INCINERATE, CRUSH, OR EXPOSE CONTENTS TO WATER.