PTC Metrology™ Temperature Calibration Laboratory
Accredited by A2LA to ISO 17025 & ANSI/NCSL Z540-1
using NIST Traceable Standards

PTC Metrology™ provides temperature calibration and certification services for Thermometers, Surface, Air, Immersion and Digital, Liquid-in-Glass, Infrared, PRT's, Thermistors, Thermocouples, and Stem Thermometers. Many quality assurance systems require that calibrated or certified thermometers be used where accurate measurements are of importance. PTC has calibrated thermometers for over 60 years and is now accredited by A2LA to ISO/IEC 17025 and ANSI/NCSL, Z540-1. Please download our scope of our accreditation from A2LA. PTC Metrology also offers calibration services for pressure, humidity and durometers. Please make your selection from the list of certifications below or contact our customer service department.

Surface Temperature

PTC Metrology™offers several levels of calibration services to conform to your quality system requirements. PTC recommends 3 point calibration as a
minimum on surface thermometers.

Temperature DevicesDigital ThermometersProbes

ThermocouplesSensorsThermistors

ThermometersPRT'sRTD's

Including Thermometers with Probes or Stems, Readouts with Sensors or Sensors Only
PRT's , RTD's, Thermistors, Thermocouples & Probes

Air Temperature

Environmental Thermometers, Data Loggers and
Recording Thermometers (without probes

PTC Models 616, 618, 815 and 818 series.

Black BodyInfrared

Infrared temperature calibration from 20°C to1000°C Black Body up to1200°C

Liquid In Glass

Including ASTM Liquid In Glass Thermometers

The mean temperature on Mars is −46 °C (-51 °F)

Tech Note on ASTM Thermometers

PTC Metrology™ will examine and calibrate your ASTM Thermometer against NIST traceable standards at temperature points specified by ASTM E1 Table 1. The thermometers will be calibrated at the specified temperature in accordance with Test Method E-77. PTC is accredited by A2LA to ISO/IEC 17025 and ANSI/NCSL Z540-1. Please contact PTC us for pricing .

Selecting Temperature Points

To assist you in choosing test points, please refer to ASTM E-77 and NIST recommendations.

1. A minimum of three temperatures should be calibrated, generally low, medium and high on the scale.

2. There should be no more than 100 graduations between any two calibration temperatures. If a thermometer is calibrated every 100 divisions, the user may interpolate intermediary corrections with a precision of ± 1/2° scale division (ASTME-77) .

3. If the temperatures used by the manufacturer for scale placement are known, the temperatures used for calibration should correspond. This will assure linearity of spacing between the calibrated points, therefore allowing the user to interpolate intermediate values.

A2LA Scope of Accreditation PTC Metrology™

Recall Notification

PTC Metrology® will notify you by email when your instrument is due for recalibration, in accordance with the period of validity that your quality system specifies. Please register your instrument today.