Features
|
Model PPS41
|
|
Two Selectable Sensors
|
One or two integral pressure ranges
|
|
Resolution
|
User selectable between 3 and 6 digits
|
|
Damping
|
User selectable up to 5 seconds
|
|
Pressure Units
|
11 Pre programmed units plus one user defined unit
|
|
Internal Memory
|
128kB
|
|
Pressure Calibration
|
Including full alphanumeric test , instrument and user id with full time and date stamp
|
|
Pressure Generation
|
Hydraulic pressure generation available in 3 ranges for improved fine control 200/400/700 bar
|
|
Audible Warning
|
For over-pressure and low battery voltage
|
|
Data Logging
|
Up to 100 hours
|
|
Software
|
Windows based automatic calibration software supplied as standard
|
|
Peak
|
Integral peak hold facility
|
|
Hold
|
Integral freeze frame facility
|
|
Tare
|
Integral tare facility for pressure or pressure and electrical values
|
|
Leak
|
The leak test function calculates leak rates over a user defined period and records the pressure change and the test temperature (with an external PT100 DIN )
|
|
Switch Test
|
Records opening and closing pressures for a switch
|
|
Backlight
|
Set to always on , off or timed auto shut off
|