Accredited by A2LA to ISO/IEC17025
ANSI/NCSL Z540-1 Certification

Specialists in Temperature Pressure RH & Durometer Calibration

PTC Metrology's calibration lab is accredited by A2LA to ISO/IEC 17025, ANSI/NCSL Z540-1 and uses NIST traceable standards. Our lab provides certification services for Durometers, Thermometers, Liquid-in-Glass, PRT's, Thermistors, Thermocouples, Pressure Gauges, Data Loggers, Hygrometers & Releative Humidity Readouts. Many quality assurance systems require that calibrated or certified instruments be used where accurate measurements are of importance. PTC has calibrated thermometers for over 55 years and now offers NIST Traceable certification accredited by A2LA to ISO 17025. Click on the link to view our scope of our accreditation . Our accreditation has been renewed until March 31, 2008.


Enter a number in either field, then click outside the text box.
F:

C:

3 Temperature Points PTC's Choosing NIST Traceable and A2LA Accredited Certification ( Air Immersion NIST only )

This certification will satisfy most quality requirements by providing NIST  traceable as received and as returned calibration results of the UUT ( unit under test) . The calibration points will be of PTC's choice and will include three points across the UUT's scale.

Customer's Selected Temperature Points NIST Traceable and A2LA Accredited Certification ( Air Immersion NIST only )

This certification will contain the as received and as returned calibration results of the UUT at the point(s) selected by the customer. Information on the customer's UUT and PTC's calibration standards, statement of uncertainty, laboratory conditions and information on the primary standards will also be on the certification. Cost of the certification will depend on which temperature points the customer chooses. To assist you in choosing temperature points, please refer to ASTME-77 and NIST recommendations. Minimum charge for customer selected temperature points ranges from $65.00 to $115.00 per instrument.

Links to PTC Metrology www.ptcmetrology.com,

Basic Certification Traceable to NIST (non A2LA)
without logged temperature points

This certification is best for customer's needing only calibration traceability to NIST without the added cost of logging the calibration results, A2LA accredited certification or meeting the requirements of ISO guide 17025. Although no specific points will be logged, UUT's referred to will be calibrated at three or more points over the UUT's range. Up to five instruments of the same model and range can be listed on the cert for the basic price. Additional units above quantity of five can be added for a nominal fee. This certification will not list any adjustments or out of tolerance conditions.

Certification price does not include labor, adjustment, repairs, or replacement parts.

Durometer Certification

PTC Metrology is accredited by A2LA to ISO/IEC 17025/ANSI/NCSL Z540-1 for Durometer calibration. (Certificate number 1896.01) All Calibration Standards are NIST Traceable.

A written certification for any durometer covered by current ASTM D2440, ASTM F1957, ISO 868, ISO 7619 and DIN 53505 specifications are available from PTC Metrology . Each certification provides "as received" and "as left" data. The points calibrated are normally every ten points over the durometer's scale.

A complete durometer calibration report must include the following:

1) indenter geometry
2) indenter extension
3) gage linearity
4) force curve

Pressure Certification

Our lab offers pressure calibration from 6 to12140 psig with an uncertainty of 50 parts in 106 with a Ruska 2400 HL Deadweight Tester. Our laboratory has the capabilities to calibrate general purpose gauges, utility, industrial and process gauges, test gauges, pressure recorders and data loggers.

Relative Humidity

PTC Metrology offers standard humidity certifications; which include as received and as returned data; a statement of uncertainty; laboratory conditions; and information on the primary standards.

Steel Hardness Block Certification

Steel hardness blocks are certified in accordance with ASTM E18. Each block is tested five times and the average value is reported, provided that their range conforms to Table 21 in ASTM E18. Certification information includes: type of standard, model number, serial number, date calibrated, and due date.

Microscope Reticle Verification NIST Traceable Certification

Dimensional verification of microscope reticle. Model 316 & 450 Hardness Scopes

Home I Products I Site Map I Shop Online I Downloads I Contact Us I PTC Metrology
PTC Instruments 2301 Federal Avenue Los Angeles Ca. 90064-1482
310-478-1134 / 877-782-2329 / Fax 310-312-0826 www.ptc1.com sales @ptc1.com
All rights reserved MMV